·Background Correction Ability: more than 60 times.
Flame Atomization System:
·The element switching device of AAS has the advantages of rapid positioning and compact structure.
·Characteristic Concentration(Cu): 0.02μg/mL/1%
·Detection Limit(Cu): 0.003μg/mL
·Precision(RSD): less than 0.6%
·Burner: high temperature resistant, corrosion-resistant metal titanium alloy; three-dimensional adjustment.
·Sprayer: high intensity mixing chamber with special treatment technology; corrosion resistance; easy cleaning; no memory effect.
·Atomization Chamber: explosion-proof; corrosion-resistant.
·Regulation System: automatic PC control and regulation (flow of gas and auxiliary gas; burner position and angle); automatic optimization; automatic ignition; automatic identification of combustion head.
·Safety Protection: a full set of safety interlocking system; various pressure monitoring; automatic safety protection functions; acetylene leakage alarm and shutdown system(automatic monitoring of combustion head, flame state, water seal, gas pressure and atomization system pressure; automatic interlocking and shutdown in case of abnormality or power failure).
·Liquid Level Detection: prohibit the overflow of waste liquid.
Graphite Furnace Atomization System:
·Temperature Range: room temperature to 3000 ºC; temperature self-correction function.
·Temperature Control Accuracy: less than 1%
·Heating Rate: more than 2000ºC/s
·Temperature Control Program: 10 heating procedures(max); 3 heating modes(step, slope and hold).
·Heating Mode: longitudinal heating
·Detection Limit(Cd): less than 0.5pg
·Precision(Cd): less than 2%
·Safety Detection Measures: graphite furnace body temperature detection; cooling water pressure detection; inertgas pressure detection; graphite tube fracture or unclamp detection.
·Safety Protection: argon underpressure indicator; cooling water flow insufficiency, overheating or overcurrent alarm; automatic protection function.
·Concentration Calculation Method: standard curve method(1~3 curves); automatic fitting; standard addition method; repeated measurement(1~99 times) and calculation(average value, standard deviation and relative standard deviation).
·Software and Data Processing: high intelligent software; automatic instrument and accessory control; automatic optimization.
·Result Printing: parameter printing; data result printing; graphic printing; report summary; test data saving and exporting in Excel for editing.